Comparing automated and manual assessments of tear break-up time using different non-invasive devices and a fluorescein procedure

Abstract

To evaluate the agreement and repeatability of an automated topography-based method for non-invasive break-up time (NIBUT) analyses in comparison with two other NIBUT procedures, the fluorescein procedure (fBUT), and with the manual assessment with the same device. In the first experiment, a semi-randomised crossover study was performed on forty-three participants (23.1 ± 2.1 years). NIBUT measurements were collected in a randomised order, in both eyes of participants with EasyTear View + (Easytear, Rovereto), Polaris, and Sirius + (CSO, Firenze). Then a fBUT was collected. The overall measurement procedure was repeated in a further session (retest) on the same day. In a second experiment, a retrospective randomised crossover study was performed on eighty-five NIBUT videos previously recorded by the Sirius+. Two observers assessed manually the videos and the NIBUTs were compared with the automatic ones. In the first experiment, ANOVA showed a significant difference between the four measures in both eyes (p < 0.001). Significant differences were found in the paired comparisons between each NIBUT procedure and fBUT (Wicoxon; p < 0.05). Sirius+ resulted in agreement only with Polaris in the left eye. Correlations between all NIBUT procedures resulted in statistical significance in both eyes. All procedures showed very good test-rest reliability. In the second experiment, a significant correlation between automated and manual NIBUT was found, but also a significant statistical difference between the two measurements, although clinically negligible (0.3 s). The investigated NIBUT devices perform differently from each other (and from fBUT), so they cannot be considered interchangeable. The automated measure of NIBUT with Sirius+ has a negligible clinical difference compared to manual assessment on the same device.

Publication DOI: https://doi.org/10.1038/s41598-024-52686-0
Divisions: College of Health & Life Sciences
Additional Information: This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Uncontrolled Keywords: Cross-Over Studies,Dry Eye Syndromes/diagnosis,Eye,Fluorescein,Humans,Lacerations,Reproducibility of Results,Retrospective Studies,Tears
Publication ISSN: 2045-2322
Last Modified: 16 Dec 2024 09:02
Date Deposited: 01 Feb 2024 09:11
Full Text Link:
Related URLs: https://www.nat ... 598-024-52686-0 (Publisher URL)
http://www.scop ... tnerID=8YFLogxK (Scopus URL)
PURE Output Type: Article
Published Date: 2024-01-30
Accepted Date: 2024-01-22
Submitted Date: 2023-10-05
Authors: Zeri, Fabrizio (ORCID Profile 0000-0003-0529-555X)
Rizzo, Giulia Carlotta
Ponzini, Erika
Tavazzi, Silvia

Download

[img]

Version: Published Version

License: Creative Commons Attribution

| Preview

Export / Share Citation


Statistics

Additional statistics for this record