Clinical outcomes after implantation of a new hydrophobic acrylic toric IOL during routine cataract surgery

Abstract

PURPOSE: To assess the clinical outcomes after implantation of a new hydrophobic acrylic toric intraocular lens (IOL) to correct preexisting corneal astigmatism in patients having routine cataract surgery. SETTING: Four hospital eye clinics throughout Europe. DESIGN: Cohort study. METHODS: This study included eyes with at least 0.75 diopter (D) of preexisting corneal astigmatism having routine cataract surgery. Phacoemulsification was performed followed by insertion and alignment of a Tecnis toric IOL. Patients were examined 4 to 8 weeks postoperatively; uncorrected distance visual acuity (UDVA), corrected distance visual acuity, manifest refraction, and keratometry were measured. Individual patient satisfaction with uncorrected vision and the surgeon’s assessment of ease of handling and performance of the IOL were also documented. The cylinder axis of the toric IOL was determined by dilated slitlamp examination. RESULTS: The study enrolled 67 eyes of 60 patients. Four to 8 weeks postoperatively, the mean UDVA was 0.15 logMAR G 0.17 (SD) and the UDVA was 20/40 or better in 88% of eyes. The mean refractive cylinder decreased significantly postoperatively, from -1.91 +/- 1.07 D to -0.67 +/- 0.54 D. No significant change in keratometric cylinder was observed. The mean absolute IOL misalignment from the intended axis was 3.4 degrees (range 0 to 12 degrees). The good UDVA resulted in high levels of patient satisfaction. CONCLUSION: Implantation of the new toric IOL was an effective, safe, and predictable method to manage corneal astigmatism in patients having routine cataract surgery.

Publication DOI: https://doi.org/10.1016/j.jcrs.2012.08.055
Divisions: College of Health & Life Sciences > School of Optometry > Optometry
College of Health & Life Sciences > School of Optometry > Optometry & Vision Science Research Group (OVSRG)
College of Health & Life Sciences
College of Health & Life Sciences > School of Optometry > Vision, Hearing and Language
Additional Information: NOTICE: this is the author’s version of a work that was accepted for publication in Journal of cataract and refractive surgery. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version was subsequently published in Sheppard, A, Wolffsohn, J, Bhatt, U, Hoffman, P, Scheider, A, Hutz, W & Shah, S, 'Clinical outcomes following implantation of a new hydrophobic acrylic toric IOL during routine cataract surgery' Journal of cataract and refractive surgery, vol. 39, no. 1 (2013) DOI http://dx.doi.org/10.1016/j.jcrs.2012.08.055
Uncontrolled Keywords: Ophthalmology,Sensory Systems,Surgery
Publication ISSN: 1873-4502
Last Modified: 19 Apr 2024 07:08
Date Deposited: 11 Mar 2019 18:26
Full Text Link:
Related URLs: http://www.scop ... tnerID=8YFLogxK (Scopus URL)
PURE Output Type: Article
Published Date: 2013-01
Authors: Sheppard, Amy (ORCID Profile 0000-0003-0035-8267)
Wolffsohn, James (ORCID Profile 0000-0003-4673-8927)
Bhatt, Uday
Hoffman, Peter
Scheider, Andreas
Hutz, Werner
Shah, Sunil (ORCID Profile 0000-0002-5373-5305)