Shah, Sunil, Laiquzzaman, Mohammed, Bhojwani, Rajan, Mantry, Sanjay and Cunliffe, Ian (2007). Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes. Investigative Ophthalmology and Visual Science, 48 (7), pp. 3026-3031.
Abstract
PURPOSE. To compare hysteresis, a novel measure of ocular rigidity (viscoelasticity) in normal and keratoconic eyes. METHODS. The study consisted of 207 normal and 93 keratoconic eyes. Eyes were diagnosed as keratoconic based on clinical examination and corneal topography. The hysteresis was measured by the Ocular Response Analyzer (ORA; Reichert Ophthalmic Instruments, Buffalo, NY). The data were recorded by Generation 3 software for the ORA. Central corneal thickness (CCT) was measured with a handheld ultrasonic pachymeter in the midpupillary axis. RESULTS. The mean hysteresis was 10.7 ± 2.0 (SD) mm Hg (range, 6.1-17.6) in normal eyes compared with 9.6 ± 2.2 mm Hg (range, 4.7-16.7) in keratoconic eyes. The difference was statistically significant (P < 0.0001, unpaired t-test). Mean CCT in the normal and keratoconic eyes was 545.0 ± 36.4 μm (range, 471-650) and 491.8 ± 54.7 μm (range, 341-611), respectively; the difference was significant (P < 0.0001, unpaired t-test). CONCLUSIONS. Hysteresis was significantly higher in normal than in keratoconic eyes. It may be a useful measurement in addition to CCT, when assessing ocular rigidity, and may be of particular importance when trying to correct intraocular measurements for increased or decreased ocular rigidity. Long-term studies of change in hysteresis may provide information on the progression of keratoconus.
Publication DOI: | https://doi.org/10.1167/iovs.04-0694 |
---|---|
Divisions: | College of Health & Life Sciences Aston University (General) |
Additional Information: | Creative Commons Attribution Non-Commercial No Derivatives License |
Uncontrolled Keywords: | Ophthalmology |
Publication ISSN: | 0146-0404 |
Last Modified: | 17 Dec 2024 08:08 |
Date Deposited: | 13 Jun 2017 13:40 |
Full Text Link: | |
Related URLs: |
http://www.scop ... tnerID=8YFLogxK
(Scopus URL) |
PURE Output Type: | Article |
Published Date: | 2007-07-01 |
Authors: |
Shah, Sunil
(
0000-0002-5373-5305)
Laiquzzaman, Mohammed Bhojwani, Rajan Mantry, Sanjay Cunliffe, Ian |
Download
Version: Published Version
License: Creative Commons Attribution Non-commercial No Derivatives
| Preview