Heitmar, Rebekka and Kalitzeos, Angelos A. (2015). Reliability of retinal vessel calibre measurements using a retinal oximeter. BMC Ophthalmology, 15 ,
Abstract
Background: Summarised retinal vessel diameters are linked to systemic vascular pathology. Monochromatic images provide best contrast to measure vessel calibres. However, when obtaining images with a dual wavelength oximeter the red-free image can be extracted as the green channel information only which in turn will reduce the number of photographs taken at a given time. This will reduce patient exposure to the camera flash and could provide sufficient quality images to reliably measure vessel calibres. Methods: We obtained retinal images of one eye of 45 healthy participants. Central retinal arteriolar and central retinal venular equivalents (CRAE and CRVE, respectively) were measured using semi-automated software from two monochromatic images: one taken with a red-free filter and one extracted from the green channel of a dual wavelength oximetry image. Results: Participants were aged between 21 and 62 years, all were normotensive (SBP: 115 (12) mmHg; DBP: 72 (10) mmHg) and had normal intra-ocular pressures (12 (3) mmHg). Bland-Altman analysis revealed good agreement of CRAE and CRVE as obtained from both images (mean bias CRAE = 0.88; CRVE = 2.82). Conclusions: Summarised retinal vessel calibre measurements obtained from oximetry images are in good agreement to those obtained using red-free photographs.
Publication DOI: | https://doi.org/10.1186/s12886-015-0174-0 |
---|---|
Divisions: | College of Health & Life Sciences > School of Optometry > Optometry College of Health & Life Sciences > School of Optometry > Optometry & Vision Science Research Group (OVSRG) College of Health & Life Sciences College of Health & Life Sciences > School of Optometry > Vision, Hearing and Language |
Additional Information: | This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
Uncontrolled Keywords: | Ophthalmology |
Publication ISSN: | 1471-2415 |
Last Modified: | 10 Dec 2024 08:07 |
Date Deposited: | 12 Jan 2016 16:00 |
Full Text Link: |
http://bmcophth ... 2886-015-0174-0 |
Related URLs: |
http://www.scop ... tnerID=8YFLogxK
(Scopus URL) |
PURE Output Type: | Article |
Published Date: | 2015-12-24 |
Accepted Date: | 2015-12-15 |
Authors: |
Heitmar, Rebekka
(
0000-0002-7657-1788)
Kalitzeos, Angelos A. |