Does rebound tonometry probe misalignment modify intraocular pressure measurements in human eyes?

Abstract

Purpose. To examine the influence of positional misalignments on intraocular pressure (IOP) measurement with a rebound tonometer. Methods. Using the iCare rebound tonometer, IOP readings were taken from the right eye of 36 healthy subjects at the central corneal apex (CC) and compared to IOP measures using the Goldmann applanation tonometer (GAT). Using a bespoke rig, iCare IOP readings were also taken 2 mm laterally from CC, both nasally and temporally, along with angular deviations of 5 and 10 degrees, both nasally and temporally to the visual axis. Results. Mean IOP ± SD, as measured by GAT, was 14.7±2.5 mmHg versus iCare tonometer readings of 17.4±3.6 mmHg at CC, representing an iCare IOP overestimation of 2.7±2.8 mmHg (P<0.001), which increased at higher average IOPs. IOP at CC using the iCare tonometer was not significantly different to values at lateral displacements. IOP was marginally underestimated with angular deviation of the probe but only reaching significance at 10 degrees nasally. Conclusions. As shown previously, the iCare tonometer overestimates IOP compared to GAT. However, IOP measurement in normal, healthy subjects using the iCare rebound tonometer appears insensitive to misalignments. An IOP underestimation of <1 mmHg with the probe deviated 10 degrees nasally reached statistical but not clinical significance levels. © 2013 Ian G. Beasley et al.

Publication DOI: https://doi.org/10.1155/2013/791084
Divisions: College of Health & Life Sciences > School of Optometry > Optometry
College of Health & Life Sciences > School of Optometry > Optometry & Vision Science Research Group (OVSRG)
College of Health & Life Sciences
College of Health & Life Sciences > School of Optometry > Vision, Hearing and Language
Additional Information: Copyright © 2013 Ian G. Beasley et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Uncontrolled Keywords: Ophthalmology
Publication ISSN: 2090-0058
Last Modified: 08 Nov 2024 08:06
Date Deposited: 01 Oct 2013 14:09
Full Text Link: http://www.hind ... op/2013/791084/
Related URLs: http://www.scop ... tnerID=8YFLogxK (Scopus URL)
PURE Output Type: Article
Published Date: 2013-09-23
Authors: Beasley, Ian G.
Laughton, Deborah S.
Coldrick, Benjamin J. (ORCID Profile 0000-0003-0969-7062)
Drew, Thomas E.
Sallah, Marium
Davies, Leon N. (ORCID Profile 0000-0002-1554-0566)

Download

[img]

Version: Published Version

License: Creative Commons Attribution


Export / Share Citation


Statistics

Additional statistics for this record