Mobile app Aston contrast sensitivity test

Kingsnorth, Alec, Drew, Tom, Grewal, Bikramjit and Wolffsohn, James S. (2016). Mobile app Aston contrast sensitivity test. Clinical and Experimental Optometry, 99 (4), 350–355.

Abstract

BACKGROUND: Contrast detection is an important aspect of the assessment of visual function; however, clinical tests evaluate limited spatial frequencies and contrasts. This study validates the accuracy and inter-test repeatability of a swept-frequency near and distance mobile app Aston contrast sensitivity test, which overcomes this limitation compared to traditional charts. METHOD: Twenty subjects wearing their full refractive correction underwent contrast sensitivity testing on the new near application (near app), distance app, CSV-1000 and Pelli-Robson charts with full correction and with vision degraded by 0.8 and 0.2 Bangerter degradation foils. In addition repeated measures using the 0.8 occluding foil were taken. RESULTS: The mobile apps (near more than distance, p = 0.005) recorded a higher contrast sensitivity than printed tests (p < 0.001); however, all charts showed a reduction in measured contrast sensitivity with degradation (p < 0.001) and a similar decrease with increasing spatial frequency (interaction > 0.05). Although the coefficient of repeatability was lowest for the Pelli-Robson charts (0.14 log units), the mobile app charts measured more spatial frequencies, took less time and were more repeatable (near: 0.26 to 0.37 log units; distance: 0.34 to 0.39 log units) than the CSV-1000 (0.30 to 0.93 log units). The duration to complete the CSV-1000 was 124 ± 37 seconds, Pelli-Robson 78 ± 27 seconds, near app 53 ± 15 seconds and distance app 107 ± 36 seconds. CONCLUSIONS: While there were differences between charts in contrast levels measured, the new Aston near and distance apps are valid, repeatable and time-efficient method of assessing contrast sensitivity at multiple spatial frequencies.

Publication DOI: https://doi.org/10.1111/cxo.12362
Divisions: Life & Health Sciences
Life & Health Sciences > Optometry
Life & Health Sciences > Ophthalmic Research Group
Additional Information: This is the peer reviewed version of the following article: Kingsnorth, A., Drew, T., Grewal, B., & Wolffsohn, J. S. (2016). Mobile app Aston contrast sensitivity test. Clinical and Experimental Optometry, 99(4), 350–355, which has been published in final form at http://dx.doi.org/10.1111/cxo.12362. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Self-Archiving.
Uncontrolled Keywords: contrast sensitivity,CSV-1000,mobile app,Pelli–Robson chart,repeatability,validity,Ophthalmology,Optometry
Full Text Link: http://onlineli ... .12362/abstract
Related URLs: http://www.scop ... tnerID=8YFLogxK (Scopus URL)
Published Date: 2016-07
Authors: Kingsnorth, Alec
Drew, Tom
Grewal, Bikramjit
Wolffsohn, James S. ( 0000-0003-4673-8927)

Download

[img]

Version: Accepted Version

| Preview

Export / Share Citation


Statistics

Additional statistics for this record