First in-vivo trials of a fibre Bragg grating based temperature profiling system

Webb, David J.; Hathaway, M.W.; Jackson, David A.; Jones, S.; Zhang, Lin and Bennion, Ian (2000). First in-vivo trials of a fibre Bragg grating based temperature profiling system. Journal of biomedical optics, 5 (1), pp. 45-50.

Abstract

We describe the results of in-vivo trials of a portable fiber Bragg grating based temperature profile monitoring system. The probe incorporates five Bragg gratings along a single fiber and prevents the gratings from being strained. Illumination is provided by a superluminescent diode, and a miniature CCD based spectrometer is used for demultiplexing. The CCD signal is read into a portable computer through a small A/D interface; the computer then calculates the positions of the center wavelengths of the Bragg gratings, providing a resolution of 0.2°C. Tests were carried out on rabbits undergoing hyperthermia treatment of the kidney and liver via inductive heating of metallic implants and comparison was made with a commercial Fluoroptic thermometry system.

Additional Information: Webb, David J., Hathaway, M.W., Jackson, D.A., Jones, S., Zhang, Lin and Bennion, Ian, First in-vivo trials of a fibre Bragg grating based temperature profiling system. Journal of biomedical optics, 5 (1), pp. 45-50, (2000). Copyright 2000 Society of Photo-Optical Instrumentation Engineers. One print or electronic copy may be made for personal use only. Systematic reproduction and distribution, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited.
Uncontrolled Keywords: fiber Bragg grating, medical temperature sensor
Published Date: 2000-01

Download

[img]

Export / Share Citation


Statistics

Additional statistics for this record